Dolphin FIS Standard Therapy Systems

Standard Dolphin FIS Mattress (35-Inch Therapy Surface)

This is the 35-inch-wide Dolphin FIS therapy mattress, 82 inches long and 10 inches deep, the configuration that fits a standard homecare or hospital bed deck. It is not an air mattress in the ordinary sense. Joerns calls the technology Fluid Immersion Simulation, and describes what it does as automatically simulating a fluid environment, so the body is carried in what the manufacturer calls a near neutrally buoyant state rather than resting on inflating and deflating cells in sequence. Dolphin, Dolphin FIS, Fluid Immersion Simulation and AutoVector are registered trademarks of Joerns Healthcare LLC. We are an authorized Joerns dealer; we are not Joerns.

The surface is driven by the AutoVector control unit, a 10 lb box that hangs at the foot of the bed. Joerns states that the AutoVector control module monitors the support surface more than 100 times per second for patient movement or surface changes, and that the system automatically measures the specific anthropometric characteristics of the individual patient as they engage the support surface. In practice that means nobody has to dial in a setting when the person shifts, slides down the bed or is sat up for a meal. Joerns is explicit that there is no need for direct intervention or manual entry to adjust comfort settings, though a manual comfort adjustment exists and the control module re-optimizes around it.

Scale drawing

Width and rated capacity are not the same thing

Dolphin FIS surface widths at 82 inches long, with the rated weight capacity of each A scale drawing. 35 inches wide, 42 inches wide, 48 inches wide — all 82 inches long, drawn to one scale so the widths compare directly. Below the footprints a bar for each width shows its rated weight capacity: 35 inch, 500 pounds; 42 inch, 500 pounds; 48 inch, 1,000 pounds. The 35 inch and 42 inch bars are the same height as each other; the 48 inch bar is 2 times as tall. 82″ long 35″ 42″ 48″ RATED WEIGHT CAPACITY 500 lb 500 lb 1,000 lb

The same three figures, in words: every surface above is 82″ long, and only the width changes.

35″ × 82″
rated 500 lb — Standard Dolphin FIS, Low-Profile System
42″ × 82″
rated 500 lb — Wide 42-Inch
48″ × 82″
rated 1,000 lb — Extra-Wide 48-Inch

Wider across is not the same as higher rated. The 42-inch surface is 7 inches wider than the 35-inch and carries exactly the same 500 lb rating. The rating steps up once, at 48 inches, to 1,000 lb. Anyone quoting you a figure between the two is quoting something no Joerns document contains.

The marked column is the surface this page is about: 35″ wide, 82″ long, rated 500 lb.

Joerns states these as mattress capacities only. The bed frame underneath has its own rated load, and the combined weight must not exceed it — which is one of the things we check with you before anything ships.

How do you want Standard Dolphin FIS? Choose monthly rental or outright purchase.

Rent by the month

$X,XXX–$Y,YYY/mo

That is a placeholder, not a rate. Our monthly pricing is not published yet. Ask the order desk and you will get the real figure in writing before you commit to anything.

  • Billed monthly for as long as the surface is in use
  • Return it when the need ends, with no long-term commitment
  • One surface matched to the patient and the care plan

Buy outright

$XX,XXX–$YY,YYY

That is a placeholder, not a price. Our purchase pricing is not published yet. Ask the order desk for a written quote for the configuration you need.

  • One-time price, with no monthly billing
  • The surface stays with the facility or the household
  • Warranty terms are the manufacturer’s; ask us to confirm them before you order
  • Often eligible for HSA or FSA funds with a Letter of Medical Necessity; your plan administrator decides.

Models we carry

  • Joerns Healthcare Dolphin® FIS® Standard Therapy Mattress with AutoVector® Control Unit
  • Shipped by freight, with the lane quoted before you order
  • Cleaned and function-tested before it ships
  • Itemized invoices for plan and facility claims

Who this rental is for, and when it is needed

Most people reach this page with a date attached. Someone is coming home in four days with a stage 3 or stage 4 pressure injury on the sacrum, or with a flap or graft that a surgeon has told the family cannot bear load, and the hospital bed being delivered comes with an ordinary foam mattress. The discharge note may name a specific surface, or it may say something vaguer about a low air loss or air-fluidized surface, and the family is left to work out what that means and who supplies it.

Joerns describes the technology as highly effective for the prevention and healing of pressure injuries through Stage 4, as well as for treating patients with post-operative flaps and grafts, and says it may also be used for patients whose medical condition precludes turning and repositioning, or where those interventions may be contraindicated. Joerns also lists patients with spinal cord injury, with an explicit condition attached: once the acute injury has been stabilized and the patient has been cleared by a physician. Those are the manufacturer’s own indications, in the manufacturer’s own hedged wording, and Joerns closes them with the sentence that governs this whole page — that its clinical indications are guidelines, to be taken only as recommendations for consideration during individual patient assessment by the clinician.

The other common reader is a spouse or an adult child who has run out of physical capacity. Turning someone every two hours through the night is possible for a week and unsustainable for three months. A surface that reduces the pressure a person is lying on does not remove the turning schedule from the care plan, and no honest supplier will tell you it does, but it changes the conversation with the nurse about how often and how far. Ask the care team what they want, and read us their exact wording.

Situations we set this up for

  • A discharge date, a hospital bed arriving, and no therapy surface named
  • A stage 3 or stage 4 pressure injury on the trunk or pelvis
  • A post-operative flap or graft that cannot take load
  • A condition that makes turning and repositioning difficult or contraindicated
  • A stabilized spinal cord injury, with physician clearance in place
  • A caregiver who can no longer reposition alone through the night
  • A Group 2 powered surface already tried, with the care team asking what is next

Models and specifications

Figures below are the manufacturer's published specifications. We confirm the exact unit and its ratings with you before it ships.

Standard Dolphin FIS at a glance

Published specifications for Standard Dolphin FIS Mattress
Sleep surface 35″ × 82″
Surface depth 10″
Weight capacity Up to 500 lb
Therapy type Autofirm; returns to the previous setting after approximately 15 minutes
Power and battery Sealed 12 VDC rechargeable; approximately 12 hours
Alarms Not published
HSA and FSA Often eligible with a Letter of Medical Necessity; your plan administrator decides

Figures are the manufacturer’s published specifications. Anything shown as "Not published" is a figure we have not confirmed in writing; ask us and we will check the spec sheet for the exact unit before it ships.

Joerns Healthcare

Dolphin® FIS® Standard Therapy Mattress with AutoVector® Control Unit

The standard therapy mattress is 35 inches wide by 82 inches long by 10 inches deep, and it weighs 22 lb. That figure surprises people who are expecting a hospital bed to arrive. What ships is a 22 lb mattress and a 10 lb control unit, not a bed — the frame is a separate question, and if you do not already have one, the surface alone does not solve the problem.

Joerns rates the standard surface at 500 lb. The 42-inch surface carries the same 500 lb rating; only the 48-inch bariatric configuration is rated 1,000 lb. And Joerns footnotes every one of those numbers the same way: it is the mattress weight capacity only, and total weight must not exceed the bed frame manufacturer’s specified load capacity. In plain English, the frame underneath has its own limit and it is usually the lower of the two. Tell us the frame make and model, not just the person’s weight.

Control is automatic. Joerns states that the AutoVector control module monitors the support surface more than 100 times per second, and that the system uses complex algorithms and a microprocessor to precisely adjust the surface and optimize for each patient’s unique anatomical features, adjusting to weight, surface area and repositioning as they happen. An Autofirm mode firms the surface for transfers and care procedures, and Joerns states the system automatically returns to the previous setting after approximately 15 minutes, so a caregiver cannot leave someone parked on a firm surface by forgetting to switch it back. A quick CPR deflation valve is fitted to the therapy mattress.

One thing this model does not have, and we will not pretend otherwise: it has no safety cell. Joerns documents a foam safety cell on Low Profile models only, and an air-filled safety cell on V-Matt step-deck models only, in both cases as protection against bottoming out during a power failure that exceeds battery life. Neither belongs to this 10-inch standard mattress. What this mattress has instead is the rechargeable sealed 12 VDC battery in the control unit, which Joerns says provides alternate power for approximately 12 hours if the system is disconnected or the mains fail. Approximately twelve hours is not a guaranteed runtime, and beyond it the standard surface has no documented bottom-out protection. If your house loses power for a day at a time in winter, say so before you choose this configuration over the Low Profile.

Dimensions
35" (89 cm) W × 82" (208 cm) L × 10" (25 cm) D
Weight capacity
500 lb — mattress capacity only; the bed frame has its own rated limit
Mattress weight
22 lb (10 kg)
Control unit
AutoVector®, 11.5" W × 12.5" H × 6" D, 10 lb (4.5 kg)
Sensing rate
Joerns: monitors the support surface more than 100 times per second
Battery backup
Sealed 12 VDC rechargeable; approximately 12 hours
Care mode
Autofirm; returns to the previous setting after approximately 15 minutes
Emergency
Quick CPR deflation valve
Cover
Smooth nylon, low friction and low shear; breathable, moisture vapour permeable
Electrical
115/230 VAC, 50/60 Hz, 0.34/0.21 A
Operating conditions
50–95 °F (10–35 °C), 30–75% RH non-condensing, IPX2
Safety agency
UL 60601-1, CAN/CSA C22.2 No. 601.1 (North America)
Warranty
Two years mattress, two years control unit (Joerns)

What makes it work at home

It sets itself, continuously

Joerns states the AutoVector control module monitors the surface more than 100 times per second and measures the patient’s own anthropometric characteristics as they engage it. There is no cycle to choose and no weight dial to get wrong.

Approximately 12 hours of battery

A sealed 12 VDC rechargeable battery in the control unit carries the system for approximately 12 hours through a disconnection or an outage, and recharges when power returns. Joerns notes the Storage Switch must be in the Battery On position to recharge.

Autofirm that times out on its own

Firm the surface for a transfer, a dressing change or a bed bath. Joerns states the system returns to the previous setting after approximately 15 minutes, so therapy resumes whether or not anyone remembers to switch it back.

A low-shear, breathable cover

Joerns describes the therapy pad fabric as smooth nylon with low friction and low shear properties, breathable and moisture vapour permeable, allowing air to circulate beneath the pad and wicking moisture away.

CPR deflation valve

A quick deflation valve drops the surface so the person is on a firm deck if chest compressions are needed. We show the caregiver where it is and what it feels like to pull, before we leave.

A 22 lb surface, not a bed

The mattress is 22 lb and the control unit 10 lb. That makes the freight simple and the stairs manageable, but it also means this is a surface for a frame you already have or are getting separately.

Delivery and setup

A Dolphin FIS system arrives as two items: the 22 lb therapy mattress and the 10 lb AutoVector control unit, with hoses. Before it ships we want the bed frame’s make and model and the deck measurement, because a 35-inch surface on a wider frame is not a matter of preference. Joerns’ own warning is blunt: the risk of entrapment can arise when equipment is placed on bed frames that leave gaps of even a few inches between the therapy mattress and the head panel, foot panel and bed or side rails, and the equipment is not to be used when such gaps are present. That is why we ask for frame details rather than a bed size, and why we would rather change the configuration before dispatch than discover the gap on the day.

On setup the existing mattress comes off the deck and the therapy surface goes on in its place. The control unit hangs at the foot, the hoses are routed so they cannot be pinched when the head or knee section articulates, and the unit goes to a wall outlet rather than a power strip that can be switched off by a foot. The surface is allowed to fill and settle, and the caregiver is walked through four things: where the CPR valve is, what Autofirm does and that it times out on its own, what the battery covers and what it does not, and the hand check — Joerns instructs that after a comfort adjustment the caregiver should confirm by hand that the patient is not touching the mattress support platform. That check belongs to whoever is looking after the person, so it is taught at setup rather than left in a manual.

How in-home setup and pickup work

Before we arrive

  • Have the bed frame make, model and deck measurement ready — not just "a hospital bed".
  • Confirm the person’s current charted weight, and ask what the frame itself is rated for.
  • Keep a dedicated wall outlet free at the foot of the bed; avoid switched outlets and power strips.
  • Plan the hose route clear of the head and knee articulation and out of the walkway.
  • Check the room stays within 50–95 °F — that is Joerns’ operating range, not a comfort preference.
  • Have the discharge summary and any wound care orders where we can read them at setup.
  • Decide who the primary caregiver is, so one person gets the walkthrough properly.
  • Know what usually happens in a power cut here, and how long it lasts.

Choosing the right one

Choose between this and its siblings on three things: the depth you can live with, the frame you have, and what happens after the battery runs out. This standard surface is 10 inches deep, which puts the sleeping height about two inches above the Low Profile. Two inches sounds trivial until someone is transferring to a wheelchair or standing from the edge of the bed, or until you look at how far the side rails now project above the surface. If the transfer is marginal, the Low Profile is the better answer and it also carries the foam safety cell that this model does not.

The frame decides the rest. A step-deck articulating frame needs the V-Matt, which is 88 inches long and contoured for that deck; putting an 82-inch flat surface on a step-deck frame is the wrong shape and reintroduces exactly the gaps Joerns warns about. A person who is within 500 lb but cannot turn without their shoulders or hips coming to the edge needs the 42-inch width rather than more capacity — it is seven inches wider across at the same 500 lb rating. Above 500 lb, or on a bariatric frame, it is the 48-inch surface and nothing else, because 1,000 lb is the only other capacity Joerns publishes for this line.

Paying for the rental

Dolphin FIS is bought or rented privately. We do not bill Medicare, and we do not submit claims on your behalf. That is not a policy preference, it is the shape of the benefit: Medicare groups pressure-reducing support surfaces into three, and there is no code for fluid immersion simulation in any of them. Group 2 covers powered surfaces — E0277, E0371, E0372 and E0373 — and its criteria are the ones a wound clinician is most likely to be able to document: multiple stage 2 ulcers on the trunk or pelvis that have failed to improve over the past month on a comprehensive treatment program, or large or multiple stage 3 or stage 4 ulcers on the trunk or pelvis, or a myocutaneous flap or skin graft for a pressure ulcer on the trunk or pelvis within the past 60 days with a Group 2 or Group 3 surface in place immediately before a discharge in the past 30 days. Group 3 is a single code, E0194, the air-fluidized bed, and an air-fluidized bed is defined by warm filtered air forced through silicone-coated ceramic beads. A Dolphin FIS is not that machine. Joerns’ own wording is that it delivers many of the best elements of air-fluidized therapy — a comparison, not an equivalence — and nobody should tell you otherwise, including us.

Two routes do apply. A therapeutic surface prescribed for a diagnosed pressure injury is exactly the kind of dual-purpose item a Letter of Medical Necessity exists for, so an HSA or FSA claim is a conversation worth having: a mattress is not on any pre-approved eligibility list by itself, a clinician’s letter naming the condition, the item, the reason and a duration is what changes that, and your plan administrator decides, not us. We provide an itemised receipt naming the item, the date and the amount, in a form suited to substantiation. We do not draft or supply the letter; that is your clinician’s document. Separately, veterans’ facilities and health networks purchase and lease equipment through their own contracting channels rather than Medicare DME rules, and we sell and lease into that route; ask us how the paperwork runs rather than assuming it mirrors a consumer claim.

Insurance and funding guides

Urgent order

Discharge tomorrow? Tell us before you commit to a date.

Send the order desk the patient’s weight, the therapy the care team has specified and the ship-to address. We will come back to you with what is in stock, what the freight lane looks like and what can realistically arrive in time, before you promise anything to the family.

  • Weight and height of the patient
  • Therapy the care team has specified
  • Ship-to address and dock or door access
  • Date the surface has to be in the room

Standard Dolphin FIS: common questions

Is a Dolphin FIS mattress an air-fluidized bed?
No. An air-fluidized bed, Medicare code E0194, works by forcing warm filtered air through silicone-coated ceramic beads. Dolphin FIS is a different technology. Joerns says it delivers many of the best elements of air-fluidized therapy, such as three-dimensional volumetric engagement and the elimination of gradient shear forces, but that is a comparison and not an equivalence, and fluid immersion simulation has no Medicare code of its own.
How much weight does the standard 35-inch Dolphin FIS hold?
Joerns rates the standard therapy mattress at 500 lb. The 42-inch surface carries the same 500 lb rating and only the 48-inch bariatric configuration is rated 1,000 lb. Joerns footnotes every capacity figure the same way: that is the mattress capacity only, and total weight must not exceed the bed frame manufacturer’s specified load capacity, which is often the lower of the two limits.
What happens in a power cut?
Joerns states the rechargeable sealed 12 VDC battery provides alternate power to the control unit for approximately 12 hours if the system is disconnected or the mains fail, and that it recharges once power is restored. It is approximately twelve hours, not a guaranteed runtime. The standard 10-inch mattress has no safety cell; the foam safety cell belongs to Low Profile models and the air-filled one to V-Matt models, both for a power failure that exceeds battery life.
Does the caregiver have to adjust anything?
Joerns states there is no need for direct intervention or manual entry to adjust comfort settings, because the AutoVector control module monitors the support surface more than 100 times per second and re-optimizes as the patient moves. A manual comfort adjustment exists, and after any adjustment Joerns instructs the caregiver to perform a hand check to confirm the patient is not touching the mattress support platform. We teach that check at setup.
Will Medicare pay for this?
We cannot tell you that it will, and we do not bill Medicare. Fluid immersion simulation has no HCPCS code, Dolphin FIS is not an air-fluidized bed, and at least one major commercial payer, Aetna, classifies fluid immersion simulation as experimental, investigational or unproven and describes it as institutional equipment inappropriate for home use. If cost is the deciding factor, ask the care team whether a covered Group 2 powered surface is clinically appropriate first.
Does it come with a bed?
No. What ships is the 22 lb therapy mattress and the 10 lb AutoVector control unit with hoses. It goes onto a bed frame you already have or are arranging separately. We ask for the frame make, model and deck measurement before dispatch, because the surface has to fill the deck without leaving gaps at the panels and rails.
Can it go on a household bed?
It is sized for a hospital bed deck at 35 by 82 inches, and it needs a deck it fits without gaps. Joerns warns that the risk of entrapment arises where gaps of even a few inches are left between the therapy mattress and the head panel, foot panel and bed or side rails, and states the equipment is not to be used where such gaps are present. Tell us what bed the person has and we will say plainly whether it works.
How warm can the room be?
Joerns’ current user-service manual gives the operating range as 50 to 95 °F, 10 to 35 °C, with relative humidity 30 to 75 per cent non-condensing. A wider figure, 50 to 104 °F, appears in Joerns documentation as the storage and shipping range, and we do not publish it as an operating ceiling. In an unconditioned room in a hot summer, this matters.
Where did this technology come from?
Joerns states that fluid immersion simulation was initially developed for the dry transport of specially trained U.S. Navy dolphins, and that it automatically simulates a fluid environment, maintaining near normal blood flow, minimizing soft tissue distortion and optimizing tissue oxygenation. It was developed for that use, not by the Navy, and there is no military endorsement of this product or of us.

Rent or buy the Standard Dolphin FIS